Procedure for Performing ACT Controls
1. Press Power and allow instrument to power on.
Note: After powering on the instrument, one or more alert messages may display.
Read the message carefully and perform the functions necessary to evaluate and/or
clear the alert. The Home screen will display when the alerts have been successfully
managed.
2. From the Home screen touch More Options > Quality Options > Quality Control. Three options are
available. The default is Perform Unscheduled QC.
3. Touch the appropriate button and continue to follow the prompts on the screen.
4. Prepare the control solution as follows:
Note: Best practice is to reconstitute and use one level at a time.
A. Allow lyophilized plasma and diluent to stand at room temperature for 45 minutes.
B. Remove the cap and stopper from lyophilized plasma and remove the cap from the diluent vial.
C. Pour the entire contents of the diluent into the lyophilized plasma vial. Discard the empty vial.
D. Place the stopper back on the reconstituted control vial, closing the vial appropriately so that the
contents do not leak or spill.
E. Allow the vial to sit at room temperature for 1 minute.
F. Mix the contents of the vial by swirling gently for 1 minute, then inverting slowly for 30 seconds.
Note: To minimize foaming of the control sample, avoid vigorous or rapid mixing
motion.
G. Visually inspect the control vial to ensure the sample is fully reconstituted. If not, discard and
repeat from step A.
H. IMMEDIATELY:
•
Transfer the solution from the vial into the cartridge using a plastic transfer pipette or a non-
anticoagulated plastic syringe.
•
Close the cartridge
•
Insert the cartridge into the instrument.
Transferring the i-STAT ACT Control solution
A plastic transfer pipette, plastic syringe, or plastic capillary tube with no anticoagulant is recommended
to transfer i-STAT ACT Control solutions from the vial to the i-STAT ACT cartridge.
To use a transfer device:
1. Place the end of the transfer device into the bottom of the vial.
2. Slowly remove approximately 1cc of solution from the vial into the transfer device.
•
If air bubbles are continually drawn into the transfer device, or if a bubble is trapped near the tip
of the transfer device, discard the vial and the transfer device.
3. Remove transfer device from the vial and expel one or two drops from the transfer device.
4. Place the end of the transfer device into the sample well of the cartridge and fill the cartridge to the
fill mark.
5. Close cartridge and begin testing immediately.
78
i-STAT Alinity — System Operations Manual
Art: 745533-01 Rev. H
Rev. Date: 02-Nov-2022
Summary of Contents for i-STAT Alinity
Page 1: ......
Page 12: ...xii i STAT Alinity System Operations Manual Art 746982 01 Rev E Rev Date 02 Nov 2022 ...
Page 24: ...xxiv i STAT Alinity System Operations Manual Art 746983 01 Rev H Rev Date 11 Nov 2020 ...
Page 39: ......
Page 44: ...6 i STAT Alinity System Operations Manual Art 745526 01 Rev H Rev Date 04 Mar 2021 ...
Page 48: ...10 i STAT Alinity System Operations Manual Art 745527 01 Rev N Rev Date 13 Oct 2022 ...
Page 84: ...46 i STAT Alinity System Operations Manual Art 745531 01 Rev G Rev Date 04 Mar 2021 ...
Page 88: ...50 i STAT Alinity System Operations Manual Art 745529 01 Rev G Rev Date 04 Mar 2021 ...
Page 96: ...58 i STAT Alinity System Operations Manual Art 745530 01 Rev D Rev Date 16 May 2022 ...
Page 100: ...62 i STAT Alinity System Operations Manual Art 745578 01 Rev C Rev Date 04 Mar 2021 ...
Page 108: ...70 i STAT Alinity System Operations Manual Art 745525 01 Rev E Rev Date 04 Mar 2021 ...
Page 110: ...72 i STAT Alinity System Operations Manual Art 745532 01 Rev D Rev Date 04 Mar 2021 ...
Page 134: ...96 i STAT Alinity System Operations Manual Art 745524 01 Rev I Rev Date 02 Nov 2022 ...
Page 147: ......
Page 164: ...126 i STAT Alinity System Operations Manual Art 747079 01 Rev F Rev Date 28 Apr 2020 ...
Page 167: ......
Page 176: ...136 i STAT Alinity System Operations Manual Art 746298 01 Rev M Rev Date 13 Oct 2022 ...
Page 250: ...210 i STAT Alinity System Operations Manual Art 746300 01 Rev N Rev Date 02 Nov 2022 ...
Page 290: ...250 i STAT Alinity System Operations Manual Art 746300 01 Rev N Rev Date 02 Nov 2022 ...
Page 296: ...256 i STAT Alinity System Operations Manual Art 746304 01 Rev H Rev Date 29 Oct 2019 ...