1.11 - Calibration Verification (Cal Ver)
Calibration Verification is a procedure performed to confirm that the calibration of an instrument or test
system has remained stable throughout the reportable range. This procedure is also called a linearity
check.
Note: Only i
‑
STAT Calibration Verification and i
‑
STAT TriControls Calibration Verification
solutions can be used with a non-customized instrument. See the AlinIQ CWi section of
this manual for instruction on creating and assigning customization profiles to
instruments.
The following four items are reasons to perform calibration verification:
1. Validate the reportable range of a test before the test system is put into use.
The accuracy of results over the entire reportable range could be assessed by testing the same
patient samples on the new system and on a system with known accuracy and comparing results
using an acceptable difference criteria.
The target values have been determined over many lots of cartridges and results on these solutions
when compared to the target values indicate the performance of a particular lot of cartridges.
2. Verify that a change in reagent lot numbers does not affect either the reportable range or control
values.
Lot-to-lot variation over the entire reportable range for any reagent system could be assessed by
testing calibration verification solutions on old and new lots in parallel. Quality controls samples with
concentrations at decision points should always be used to assess new lots of reagent before results
are reported.
3. Verify that results have not been affected by maintenance or repair procedures.
The user cannot perform any maintenance procedures on the i
‑
STAT Alinity System. The software in
the instrument is updated periodically. Calibration verification solutions could be tested to verify that
the system performs as before the upgrade.
Repaired and newly purchased instruments are received with factory calibration. Testing calibration
verification samples or comparing patient sample results on a new or repaired instrument with an
older instrument will assess cartridge performance. The Electronic Simulator, rather than calibration
verification or control solutions, provides better assurance that the instrument is functioning
correctly. Any variations in instrument performance will not be statistically discernable above the
performance of the cartridges. When multiple instruments are to be used at a facility, Abbott Point of
Care Inc. recommends including at least two instruments in any performance verification studies so
that statistics reflect the “system.”
4. Troubleshoot when control values are out-of-range.
Should quality control sample results fall outside of the acceptable ranges, the use of calibration
verification samples with very low or very high concentrations could be helpful in characterizing a
reagent problem. The characteristics of the sensors and results of control solutions are sufficient for
Technical Support specialists to help users resolve out-of-range control problems.
79
i-STAT Alinity — System Operations Manual
Art: 745537-01 Rev. F
Rev. Date: 02-Nov-2022
Summary of Contents for i-STAT Alinity
Page 1: ......
Page 12: ...xii i STAT Alinity System Operations Manual Art 746982 01 Rev E Rev Date 02 Nov 2022 ...
Page 24: ...xxiv i STAT Alinity System Operations Manual Art 746983 01 Rev H Rev Date 11 Nov 2020 ...
Page 39: ......
Page 44: ...6 i STAT Alinity System Operations Manual Art 745526 01 Rev H Rev Date 04 Mar 2021 ...
Page 48: ...10 i STAT Alinity System Operations Manual Art 745527 01 Rev N Rev Date 13 Oct 2022 ...
Page 84: ...46 i STAT Alinity System Operations Manual Art 745531 01 Rev G Rev Date 04 Mar 2021 ...
Page 88: ...50 i STAT Alinity System Operations Manual Art 745529 01 Rev G Rev Date 04 Mar 2021 ...
Page 96: ...58 i STAT Alinity System Operations Manual Art 745530 01 Rev D Rev Date 16 May 2022 ...
Page 100: ...62 i STAT Alinity System Operations Manual Art 745578 01 Rev C Rev Date 04 Mar 2021 ...
Page 108: ...70 i STAT Alinity System Operations Manual Art 745525 01 Rev E Rev Date 04 Mar 2021 ...
Page 110: ...72 i STAT Alinity System Operations Manual Art 745532 01 Rev D Rev Date 04 Mar 2021 ...
Page 134: ...96 i STAT Alinity System Operations Manual Art 745524 01 Rev I Rev Date 02 Nov 2022 ...
Page 147: ......
Page 164: ...126 i STAT Alinity System Operations Manual Art 747079 01 Rev F Rev Date 28 Apr 2020 ...
Page 167: ......
Page 176: ...136 i STAT Alinity System Operations Manual Art 746298 01 Rev M Rev Date 13 Oct 2022 ...
Page 250: ...210 i STAT Alinity System Operations Manual Art 746300 01 Rev N Rev Date 02 Nov 2022 ...
Page 290: ...250 i STAT Alinity System Operations Manual Art 746300 01 Rev N Rev Date 02 Nov 2022 ...
Page 296: ...256 i STAT Alinity System Operations Manual Art 746304 01 Rev H Rev Date 29 Oct 2019 ...