Software Update and Installation via Abbott Managed Server
or the AlinIQ SDi, and a Wireless Network Connection
Use this procedure to update the instrument software using a wireless connection to Abbott’s Managed
Server or the SDi.
Note:
•
The healthcare organization network must allow instruments to connect to the Abbott Managed
Server or an SDi.
•
This procedure may take between 15 minutes and 1 hour to complete. Therefore, it is recommended
this procedure be executed outside of the clinical work area.
•
See the AlinIQ SDi User Guide for information on use of the SDi to facilitate software delivery to
instruments.
Prerequisites:
Equipment:
•
i
‑
STAT Alinity instrument(s) with appropriate Network Configuration file installed
•
Wireless network
•
Internet access
Before attempting to update instrument software from the Abbott Managed Server or the SDi, ensure
the following:
1. The instrument is configured with a network configuration that allows wireless connection to the
internet.
AlinIQ NCi - Network Connectivity for i
for instructions on configuring instrument for
network connection. The wireless connection must be enabled. If the healthcare organization
network requires use of a proxy to access the internet, the Proxy Server settings must also be
configured.
2. The instrument has a wireless connection to the healthcare organization network.
To verify:
A. Home screen shows
symbol in the upper right-hand corner and the wireless signal
strength is 3 solid bars or greater.
B. From the Home screen, navigate to: More Options > Instrument Options > Network Settings >
Network Status
C. Locate the Wireless Connection Details section and verify the wireless network interface has an IP
address.
3. The instrument is running software version OSi05 or later.
To verify:
A. From the Home screen, navigate to: More Options > Instrument Status
B. Locate the Firmware item and verify the name of the firmware is OSi05 or later (i.e., OSi06, OSi07,
OSi08, etc.).
4. Network Services for software update enabled on instrument.
To verify:
A. From the Home screen, navigate to: More Options > Instrument Options > Network Settings >
Network Services
105
i-STAT Alinity — System Operations Manual
Art: 754496-01 Rev. E
Rev. Date: 02-Nov-2022
Summary of Contents for i-STAT Alinity
Page 1: ......
Page 12: ...xii i STAT Alinity System Operations Manual Art 746982 01 Rev E Rev Date 02 Nov 2022 ...
Page 24: ...xxiv i STAT Alinity System Operations Manual Art 746983 01 Rev H Rev Date 11 Nov 2020 ...
Page 39: ......
Page 44: ...6 i STAT Alinity System Operations Manual Art 745526 01 Rev H Rev Date 04 Mar 2021 ...
Page 48: ...10 i STAT Alinity System Operations Manual Art 745527 01 Rev N Rev Date 13 Oct 2022 ...
Page 84: ...46 i STAT Alinity System Operations Manual Art 745531 01 Rev G Rev Date 04 Mar 2021 ...
Page 88: ...50 i STAT Alinity System Operations Manual Art 745529 01 Rev G Rev Date 04 Mar 2021 ...
Page 96: ...58 i STAT Alinity System Operations Manual Art 745530 01 Rev D Rev Date 16 May 2022 ...
Page 100: ...62 i STAT Alinity System Operations Manual Art 745578 01 Rev C Rev Date 04 Mar 2021 ...
Page 108: ...70 i STAT Alinity System Operations Manual Art 745525 01 Rev E Rev Date 04 Mar 2021 ...
Page 110: ...72 i STAT Alinity System Operations Manual Art 745532 01 Rev D Rev Date 04 Mar 2021 ...
Page 134: ...96 i STAT Alinity System Operations Manual Art 745524 01 Rev I Rev Date 02 Nov 2022 ...
Page 147: ......
Page 164: ...126 i STAT Alinity System Operations Manual Art 747079 01 Rev F Rev Date 28 Apr 2020 ...
Page 167: ......
Page 176: ...136 i STAT Alinity System Operations Manual Art 746298 01 Rev M Rev Date 13 Oct 2022 ...
Page 250: ...210 i STAT Alinity System Operations Manual Art 746300 01 Rev N Rev Date 02 Nov 2022 ...
Page 290: ...250 i STAT Alinity System Operations Manual Art 746300 01 Rev N Rev Date 02 Nov 2022 ...
Page 296: ...256 i STAT Alinity System Operations Manual Art 746304 01 Rev H Rev Date 29 Oct 2019 ...