
Electronic value assignment sheets (eVAS or ReVAS) may be downloaded to a customized instrument via
the Abbott Managed Server, the SDi, or USB. See the System Operations Manual for information and
instruction on creating customization profiles. Selection of eVAS or ReVAS must be made during the
creation of the profile.
Note: Follow facility policy regarding control results that do not fall within assigned
ranges.
Disposal
Check with authorities for local, state, and/or national requirements for disposal.
Procedure for Testing
Prerequisites
•
Ampules, cartridges, and instruments must be at the same temperature.
•
Controls solutions require different temperature stabilization times depending on whether PO
2
is to
be measured. If PO
2
is to be measured, equilibrate the ampule for 4 hours. If not, equilibrate the
ampule for approximately 30 minutes at room (ambient) temperature.
•
Do not use the solution left in a syringe, ampule or capillary tube for additional testing of cartridges
that contain sensors for iCa, pH, PCO
2
or PO
2
. However, cartridges without these sensors may be
tested with remaining fluids if within 10 minutes of opening the ampule.
•
Since aqueous based solutions such as controls lack the buffering capabilities of whole blood, the
transfer process from ampule to cartridge must be more expedient than with a patient sample.
Performing Quality Control Testing
1. Press Power and allow instrument to power on.
Note: After powering on the instrument, one or more alert messages may display.
Read the message carefully and perform the functions necessary to evaluate and/or
clear the alert. The Home screen will display when the alerts have been successfully
managed.
2. From the Home screen touch More Options > Quality Options > Quality Control. Select the button
that is appropriate for the testing. Continue to follow the prompts on the screen. There are three
pathways available for performing liquid QC tests. The default is unscheduled. The other two options
are available when the instrument is customized by the system administrator (see the Customization
Workspace for i
‑
STAT (AlinIQ CWi)).
3. Enter the required information by following the prompts on the screen. Prepare the cartridge and
control solution for testing.
4. Shake the ampule vigorously for 5 to 10 seconds to equilibrate the liquid and gas phases. To shake,
hold the ampule at the tip and bottom with forefinger and thumb to minimize increasing the
temperature of the solution. If necessary, tap the tip of the ampule to send solution back into the
bottom section of the ampule.
5. Protect fingers with gauze, tissue or glove, or use an ampule breaker to snap off the tip of the ampule
at the neck.
6. Immediately transfer the solution from the ampule into a capillary tube or syringe (refer to
information for transferring the control solution, below), and then immediately transfer the solution
into a cartridge. Help graphics on the instrument show the transfer of control material to the
cartridge via a syringe with a blunt tip needle.
74
i-STAT Alinity — System Operations Manual
Art: 745533-01 Rev. H
Rev. Date: 02-Nov-2022
Summary of Contents for i-STAT Alinity
Page 1: ......
Page 12: ...xii i STAT Alinity System Operations Manual Art 746982 01 Rev E Rev Date 02 Nov 2022 ...
Page 24: ...xxiv i STAT Alinity System Operations Manual Art 746983 01 Rev H Rev Date 11 Nov 2020 ...
Page 39: ......
Page 44: ...6 i STAT Alinity System Operations Manual Art 745526 01 Rev H Rev Date 04 Mar 2021 ...
Page 48: ...10 i STAT Alinity System Operations Manual Art 745527 01 Rev N Rev Date 13 Oct 2022 ...
Page 84: ...46 i STAT Alinity System Operations Manual Art 745531 01 Rev G Rev Date 04 Mar 2021 ...
Page 88: ...50 i STAT Alinity System Operations Manual Art 745529 01 Rev G Rev Date 04 Mar 2021 ...
Page 96: ...58 i STAT Alinity System Operations Manual Art 745530 01 Rev D Rev Date 16 May 2022 ...
Page 100: ...62 i STAT Alinity System Operations Manual Art 745578 01 Rev C Rev Date 04 Mar 2021 ...
Page 108: ...70 i STAT Alinity System Operations Manual Art 745525 01 Rev E Rev Date 04 Mar 2021 ...
Page 110: ...72 i STAT Alinity System Operations Manual Art 745532 01 Rev D Rev Date 04 Mar 2021 ...
Page 134: ...96 i STAT Alinity System Operations Manual Art 745524 01 Rev I Rev Date 02 Nov 2022 ...
Page 147: ......
Page 164: ...126 i STAT Alinity System Operations Manual Art 747079 01 Rev F Rev Date 28 Apr 2020 ...
Page 167: ......
Page 176: ...136 i STAT Alinity System Operations Manual Art 746298 01 Rev M Rev Date 13 Oct 2022 ...
Page 250: ...210 i STAT Alinity System Operations Manual Art 746300 01 Rev N Rev Date 02 Nov 2022 ...
Page 290: ...250 i STAT Alinity System Operations Manual Art 746300 01 Rev N Rev Date 02 Nov 2022 ...
Page 296: ...256 i STAT Alinity System Operations Manual Art 746304 01 Rev H Rev Date 29 Oct 2019 ...