![background image](http://html1.mh-extra.com/html/abbott/i-stat-alinity/i-stat-alinity_operation-manual_3045656294.webp)
Issue
Solution
Instrument time does not agree with data
manager time
1. Power on the instrument and allow it sufficient
time on the Home screen to connect to the
network and communicate with the data
manager.
2. Check the network status icon on the
instrument screen. If icon appears but
indicates no connection:
For Wired connection:
- Ensure instrument is properly seated in
the base station.
- Confirm the base station has a wired
connection to the network.
For Wireless connection:
- Ensure Wireless is Enabled via the Enable/
Disable Wireless workflow.
- Check Wireless Communication is Enabled
in the customization profile, General
Settings Category. Follow CWi instructions
in this manual to make necessary changes.
3. Check that the network status icon on the
instrument screen shows instrument has a
network connection. If not, then use NCi to
verify the NC settings within the NC file
installed on the instrument are correct. Follow
NCi instructions in this manual to make
necessary changes to network settings.
4. Check Connectivity Map in customization
profile to confirm at least one of the following
is Enabled and has the correct IP Address and
Port of the data manager: Test Results,
Operator List, Cartridge List, Patient List. Follow
CWi instructions in this manual to make
necessary changes.
5. Check the data manager to ensure it is
operational and supports communication with
i
‑
STAT Alinity instruments.
Additional Information
AlinIQ CWi is a web-based application and therefore may experience periods of downtime. Scheduled
downtime will be noted when logging in. Unscheduled downtime, although never desired, may happen
from time to time. (For example: security patches and application updates.)
Best practice would be to save all assigned profiles in a file on a healthcare organization manager's
computer. If a profile needs to be loaded onto an instrument while the system is down, upload can be
254
i-STAT Alinity — System Operations Manual
Art: 746304-01 Rev. H
Rev. Date: 29-Oct-2019
Summary of Contents for i-STAT Alinity
Page 1: ......
Page 12: ...xii i STAT Alinity System Operations Manual Art 746982 01 Rev E Rev Date 02 Nov 2022 ...
Page 24: ...xxiv i STAT Alinity System Operations Manual Art 746983 01 Rev H Rev Date 11 Nov 2020 ...
Page 39: ......
Page 44: ...6 i STAT Alinity System Operations Manual Art 745526 01 Rev H Rev Date 04 Mar 2021 ...
Page 48: ...10 i STAT Alinity System Operations Manual Art 745527 01 Rev N Rev Date 13 Oct 2022 ...
Page 84: ...46 i STAT Alinity System Operations Manual Art 745531 01 Rev G Rev Date 04 Mar 2021 ...
Page 88: ...50 i STAT Alinity System Operations Manual Art 745529 01 Rev G Rev Date 04 Mar 2021 ...
Page 96: ...58 i STAT Alinity System Operations Manual Art 745530 01 Rev D Rev Date 16 May 2022 ...
Page 100: ...62 i STAT Alinity System Operations Manual Art 745578 01 Rev C Rev Date 04 Mar 2021 ...
Page 108: ...70 i STAT Alinity System Operations Manual Art 745525 01 Rev E Rev Date 04 Mar 2021 ...
Page 110: ...72 i STAT Alinity System Operations Manual Art 745532 01 Rev D Rev Date 04 Mar 2021 ...
Page 134: ...96 i STAT Alinity System Operations Manual Art 745524 01 Rev I Rev Date 02 Nov 2022 ...
Page 147: ......
Page 164: ...126 i STAT Alinity System Operations Manual Art 747079 01 Rev F Rev Date 28 Apr 2020 ...
Page 167: ......
Page 176: ...136 i STAT Alinity System Operations Manual Art 746298 01 Rev M Rev Date 13 Oct 2022 ...
Page 250: ...210 i STAT Alinity System Operations Manual Art 746300 01 Rev N Rev Date 02 Nov 2022 ...
Page 290: ...250 i STAT Alinity System Operations Manual Art 746300 01 Rev N Rev Date 02 Nov 2022 ...
Page 296: ...256 i STAT Alinity System Operations Manual Art 746304 01 Rev H Rev Date 29 Oct 2019 ...