check failure code consisting of a numeric code, a cause and resolution message, and suggested
corrective action.
Although a message is displayed on the instrument when a quality check failure occurs, the details are
not available in Review Results. The information shown below is to aid the administrator in answering
questions or performing troubleshooting.
A Quality Check Failure can occur during patient testing (Patient Pathway), while performing Quality
Control (Quality Control Pathway), or when using the Electronic Simulator (Electronic Simulator
Pathway).
For instrument issues, follow the instructions on the display. If the instrument has been powered down,
the Quality Check Failure code will be stored in Review Results. To retrieve it: Power on the instrument
and touch More Options > Review Results > Quality Results > Quality Check Code Results.
Use the following tables to find the Quality Check Failure code and determine the cause and resolution.
1. In the table below, in the first column, find the QCF code as found in Review Results.
2. Identify the pathway in which the failure occurred, and
•
in the Cause column find the cause number, then see
for the
description
•
in the Resolution column find the resolution letter, then see
Quality Check Failure Resolutions
for
corrective action.
Quality Check Failure Codes
Note: An empty cell indicates that there is no applicable information for that pathway
and code.
QCF Code
shown in
Review Results
Patient Pathway
Quality Control
Pathway
Electronic Simulator/
Conditioning Cartridge
Pathway
QCF
Code
shown
on
screen
Cause
Resolution
Cause
Resolution
Cause
Resolution
2-01-1.1.1
1
A
1
A
2-01
2-02-1.1.2
1
A
1
A
2-02
8-01-2.1.8
I
I
P / BB
8-01
11-01-2.13.32
B
B
D
11-01
13-01-1.6.1
S
S
S
13-01
17-01-8.2.1
2
E
2
F
17-01
18-01-8.2.2
2
E
2
F
18-01
19-01-8.2.3
2
E
2
F
19-01
20-01-3.1.1
1
G
F
20-01
21-01-3.1.3
3
G
3
F
21-01
22-01-6.1.3
4
X
4
F
22-01
22-01-6.1.5
4
X
4
F
22-01
30
i-STAT Alinity — System Operations Manual
Art: 745527-01 Rev. N
Rev. Date: 13-Oct-2022
Summary of Contents for i-STAT Alinity
Page 1: ......
Page 12: ...xii i STAT Alinity System Operations Manual Art 746982 01 Rev E Rev Date 02 Nov 2022 ...
Page 24: ...xxiv i STAT Alinity System Operations Manual Art 746983 01 Rev H Rev Date 11 Nov 2020 ...
Page 39: ......
Page 44: ...6 i STAT Alinity System Operations Manual Art 745526 01 Rev H Rev Date 04 Mar 2021 ...
Page 48: ...10 i STAT Alinity System Operations Manual Art 745527 01 Rev N Rev Date 13 Oct 2022 ...
Page 84: ...46 i STAT Alinity System Operations Manual Art 745531 01 Rev G Rev Date 04 Mar 2021 ...
Page 88: ...50 i STAT Alinity System Operations Manual Art 745529 01 Rev G Rev Date 04 Mar 2021 ...
Page 96: ...58 i STAT Alinity System Operations Manual Art 745530 01 Rev D Rev Date 16 May 2022 ...
Page 100: ...62 i STAT Alinity System Operations Manual Art 745578 01 Rev C Rev Date 04 Mar 2021 ...
Page 108: ...70 i STAT Alinity System Operations Manual Art 745525 01 Rev E Rev Date 04 Mar 2021 ...
Page 110: ...72 i STAT Alinity System Operations Manual Art 745532 01 Rev D Rev Date 04 Mar 2021 ...
Page 134: ...96 i STAT Alinity System Operations Manual Art 745524 01 Rev I Rev Date 02 Nov 2022 ...
Page 147: ......
Page 164: ...126 i STAT Alinity System Operations Manual Art 747079 01 Rev F Rev Date 28 Apr 2020 ...
Page 167: ......
Page 176: ...136 i STAT Alinity System Operations Manual Art 746298 01 Rev M Rev Date 13 Oct 2022 ...
Page 250: ...210 i STAT Alinity System Operations Manual Art 746300 01 Rev N Rev Date 02 Nov 2022 ...
Page 290: ...250 i STAT Alinity System Operations Manual Art 746300 01 Rev N Rev Date 02 Nov 2022 ...
Page 296: ...256 i STAT Alinity System Operations Manual Art 746304 01 Rev H Rev Date 29 Oct 2019 ...