Closure
The closure creates an airtight seal necessary for proper fluid movement within the cartridge. The
closure also ensures that calibrant and sample remain contained within the cartridge during the testing
cycle and subsequent disposal.
Cartridge on the left in the illustration contains natural rubber latex on the snap closure. The cartridge
shown on the right is not made with natural rubber latex.
Heating elements
All i
‑
STAT cartridges require thermal control at 37°C (98.6°F), and include heating elements on the
underside of the sensor chips which are contacted and heated by the instrument's thermal probes.
Standardization and calibration
Standardization is the process by which a manufacturer establishes “true” values for representative
samples. A multi-point calibration curve, the slope or sensitivity of which is defined by coefficients in the
CLEW software, is derived for each sensor by this standardization process. These calibration curves are
stable over many lots.
A one-point calibration is performed each time a cartridge requiring calibration is used. During the first
part of the testing cycle, the calibrant solution is automatically released from its foil pack and is
positioned over the sensors. The signals produced by the sensors’ responses to the calibrant solution are
measured. This one-point calibration adjusts the offset of the stored calibration curve. Next, the
instrument automatically moves the sample over the sensors and the signals produced by the sensors’
responses to the sample are measured. While coefficients are used rather than graphic calibration
curves, the calculation of the result is equivalent to reading the sample’s concentration from an adjusted
calibration curve.
Types of cartridge sensors
Sensors are thin film electrodes microfabricated onto silicon chips. Sensing functionality is imparted to
each electrode by a number of chemically sensitive films coated over the active region of the electrodes.
The cartridges have three different types of sensors built in: potentiometric, amperometric, and
conductometric.
Potentiometric sensors
In potentiometric measurements, the difference in potential that exists between an indicator
electrode and a reference electrode is measured. Ion-selective electrodes (ISE) are examples of
potentiometric sensors. The indicator electrode is designed to be sensitive to a particular ion in a
solution. In cases where other ions are sensed by the system, selectivity coefficients can be used
to correct for this interference. An enzyme can be added to an ISE to produce ions from analytes
of interest that are not themselves ions.
Potentiometric sensors utilize two important concepts. The first concept is the Nernst Equation
which relates the measured potential to the activity of the ion being measured. It is written as:
E = E° + RT/nF ln a
91
i-STAT Alinity — System Operations Manual
Art: 745524-01 Rev. I
Rev. Date: 02-Nov-2022
Summary of Contents for i-STAT Alinity
Page 1: ......
Page 12: ...xii i STAT Alinity System Operations Manual Art 746982 01 Rev E Rev Date 02 Nov 2022 ...
Page 24: ...xxiv i STAT Alinity System Operations Manual Art 746983 01 Rev H Rev Date 11 Nov 2020 ...
Page 39: ......
Page 44: ...6 i STAT Alinity System Operations Manual Art 745526 01 Rev H Rev Date 04 Mar 2021 ...
Page 48: ...10 i STAT Alinity System Operations Manual Art 745527 01 Rev N Rev Date 13 Oct 2022 ...
Page 84: ...46 i STAT Alinity System Operations Manual Art 745531 01 Rev G Rev Date 04 Mar 2021 ...
Page 88: ...50 i STAT Alinity System Operations Manual Art 745529 01 Rev G Rev Date 04 Mar 2021 ...
Page 96: ...58 i STAT Alinity System Operations Manual Art 745530 01 Rev D Rev Date 16 May 2022 ...
Page 100: ...62 i STAT Alinity System Operations Manual Art 745578 01 Rev C Rev Date 04 Mar 2021 ...
Page 108: ...70 i STAT Alinity System Operations Manual Art 745525 01 Rev E Rev Date 04 Mar 2021 ...
Page 110: ...72 i STAT Alinity System Operations Manual Art 745532 01 Rev D Rev Date 04 Mar 2021 ...
Page 134: ...96 i STAT Alinity System Operations Manual Art 745524 01 Rev I Rev Date 02 Nov 2022 ...
Page 147: ......
Page 164: ...126 i STAT Alinity System Operations Manual Art 747079 01 Rev F Rev Date 28 Apr 2020 ...
Page 167: ......
Page 176: ...136 i STAT Alinity System Operations Manual Art 746298 01 Rev M Rev Date 13 Oct 2022 ...
Page 250: ...210 i STAT Alinity System Operations Manual Art 746300 01 Rev N Rev Date 02 Nov 2022 ...
Page 290: ...250 i STAT Alinity System Operations Manual Art 746300 01 Rev N Rev Date 02 Nov 2022 ...
Page 296: ...256 i STAT Alinity System Operations Manual Art 746304 01 Rev H Rev Date 29 Oct 2019 ...