Calibration Verification testing and i-STAT TriControls
Calibration Verification Solutions
A five-level calibration verification set is available to verify the calibration of i
‑
STAT cartridges throughout
the reportable range.
i-STAT Calibration Verification (Cal Ver) Solutions
Sodium
PCO
2
Glucose
Potassium
PO
2
Lactate
Chloride
pH
BUN/Urea
Ionized Calcium
Creatinine
i-STAT TriControls Calibration Verification Solutions
Sodium
PCO
2
Glucose
Potassium
PO
2
Lactate
Chloride
TCO
2
BUN/Urea
Ionized Calcium
Hematocrit
Creatinine
pH
Each set contains four 1.7 mL glass ampules of each level.
Storage
Store refrigerated at 2 to 8°C (35 to 46°F) until the printed expiration date on the box and ampule labels.
Unopened ampules of i
‑
STAT Calibration Verification and i
‑
STAT TriControls Calibration Verification and
solutions can also be maintained at room temperature (18 to 30°C or 64 to 86°F ) for up to 5 days.
Do not use i
‑
STAT Calibration Verification and i
‑
STAT TriControls Calibration Verification solutions past
the labeled expiration date on the box and ampule labels.
Ampule use
Barcode on the ampule must be scanned before the ampule is opened.
i
‑
STAT Calibration Verification and i
‑
STAT TriControls Calibration Verification ampule testing must be
performed within 10 minutes of opening the ampule.
i
‑
STAT Calibration Verification and i
‑
STAT TriControls Calibration Verification ampule testing must be
performed immediately upon opening the ampule when testing pH, PCO
2
, PO
2
, iCa.
Ranges
Assigned ranges are provided in the form of value assignment sheets (VAS). These are located on
www.globalpointofcare.abbott/en/support/istat-brand/istat-alinity.html
. Abbott Point of Care makes
80
i-STAT Alinity — System Operations Manual
Art: 745537-01 Rev. F
Rev. Date: 02-Nov-2022
Summary of Contents for i-STAT Alinity
Page 1: ......
Page 12: ...xii i STAT Alinity System Operations Manual Art 746982 01 Rev E Rev Date 02 Nov 2022 ...
Page 24: ...xxiv i STAT Alinity System Operations Manual Art 746983 01 Rev H Rev Date 11 Nov 2020 ...
Page 39: ......
Page 44: ...6 i STAT Alinity System Operations Manual Art 745526 01 Rev H Rev Date 04 Mar 2021 ...
Page 48: ...10 i STAT Alinity System Operations Manual Art 745527 01 Rev N Rev Date 13 Oct 2022 ...
Page 84: ...46 i STAT Alinity System Operations Manual Art 745531 01 Rev G Rev Date 04 Mar 2021 ...
Page 88: ...50 i STAT Alinity System Operations Manual Art 745529 01 Rev G Rev Date 04 Mar 2021 ...
Page 96: ...58 i STAT Alinity System Operations Manual Art 745530 01 Rev D Rev Date 16 May 2022 ...
Page 100: ...62 i STAT Alinity System Operations Manual Art 745578 01 Rev C Rev Date 04 Mar 2021 ...
Page 108: ...70 i STAT Alinity System Operations Manual Art 745525 01 Rev E Rev Date 04 Mar 2021 ...
Page 110: ...72 i STAT Alinity System Operations Manual Art 745532 01 Rev D Rev Date 04 Mar 2021 ...
Page 134: ...96 i STAT Alinity System Operations Manual Art 745524 01 Rev I Rev Date 02 Nov 2022 ...
Page 147: ......
Page 164: ...126 i STAT Alinity System Operations Manual Art 747079 01 Rev F Rev Date 28 Apr 2020 ...
Page 167: ......
Page 176: ...136 i STAT Alinity System Operations Manual Art 746298 01 Rev M Rev Date 13 Oct 2022 ...
Page 250: ...210 i STAT Alinity System Operations Manual Art 746300 01 Rev N Rev Date 02 Nov 2022 ...
Page 290: ...250 i STAT Alinity System Operations Manual Art 746300 01 Rev N Rev Date 02 Nov 2022 ...
Page 296: ...256 i STAT Alinity System Operations Manual Art 746304 01 Rev H Rev Date 29 Oct 2019 ...