2-3
Questions? Refer to Page 11-22 For Technical Support Instructions.
Data Entry Keypad
The data entry keypad allows you to enter identification numbers and letters (only the 26 letter
English alphabet) or to select an option that appears on the display. The keypad contains a 10 digit
telephone-style keypad with the keys for the numbers 0-9. The 2-9 keys also have letters printed
below the number. The keypad also has six special keys, including a Scan button, On/Off button,
Backlight/Alphanumeric, Clear, Menu and Enter keys.
Scan
Press this button to operate the barcode scanner. If the audible indicator is enabled, the meter
will beep once a barcode has been successfully scanned. Continuing to hold the Scan button
will display the scanned barcode on the display until the button is released. If no information is
scanned after 3 seconds, the scanner will turn off.
On/Off
Press this button to turn the meter on. Each time the meter is turned on, the Abbott logo screen will appear in black for a few
seconds and then will appear in grey. This provides an opportunity to ensure that all the pixels are working correctly.
Press and hold this button for 2 seconds to turn the meter off. The meter will automatically shut itself off after 4 minutes to
conserve battery life. The length of time for automatic shut-off can be configured from 4 to 10 minutes following patient tests
only.
Chapter 2: Using the Meter
Summary of Contents for FreeStyle Precision Pro
Page 1: ...Operator s Manual for Healthcare Professionals...
Page 88: ...6 Proficiency Test Proficiency Test Glucose Proficiency Test Ketone Beta Hydroxybutyrate...
Page 98: ...7 Linearity Test...
Page 135: ...9 Review Setup for Wireless Communication Optional Meter Wireless Configuration...
Page 179: ...13 Specifications FreeStyle Precision Pro Meter Symbols Barcode Types...