11-17
Questions? Refer to Page 11-22 For Technical Support Instructions.
Symptom/Error Message
Explanation
How To Respond
The Operator ID entered is not on the list
of acceptable Operator IDs defined for this
meter.
1.
Press
1
to
ReEnter
the Operator ID.
2.
If the problem persists, contact the
system administrator for further
information on operator certification.
The Operator ID entered has expired.
(The menu items
1 - ReEnter ID
and
2
-
Continue
appear only if this option is
set to warn.)
1.
Press
1
to
ReEnter
the ID using the
barcode scanner or keypad.
Or,
Press
2
to
Continue
testing.
2.
If the problem persists, notify the
system administrator.
The Operator ID is due to expire. At this
time, testing is still allowed.
1.
Press
1
to
Continue
with testing.
2.
Contact your manager or the system
administrator for further information
about operator certification.
Chapter 11: Troubleshooting