Chapter 1: Overview
Control Solutions
Control solutions are used to perform regular quality control checks on the meter to ensure it is functioning correctly. Control
testing is further described in Chapter 4. Use MediSense Glucose and Ketone Control Solutions with the FreeStyle Precision Pro
system. These control solutions are available in one, two or three level configurations.
Calibration Verification Controls
Use RNA Medical brand Glucose and β-Ketone Calibration Verification Controls to confirm the calibration and analytical
measurement range of the meter for FreeStyle Precision Pro Blood Glucose or Blood β-Ketone Test Strips. This is called linearity
testing and requires a kit with 5 levels of control solution. Linearity testing is further described in Chapter 7.
Replacement Strip Port Module
A replacement strip port module is available if the system becomes disabled as a result of blood, fluid or other contaminant
entering the strip port. Repeated strip port error messages may also indicate the need for replacement. The strip port module
replacement procedure is described in Chapter 12.
Data Management System
Abbott’s FreeStyle PrecisionWeb Point-of-Care Health Management System (PrecisionWeb) is necessary to upload results from
the FreeStyle Precision Pro meter. Data upload from remote locations requires either terminal servers or networked PCs running
Abbott’s Data Repeater software. Data can also be uploaded through a wireless network when the individual meters and data
management system are properly enabled. Wireless configuration details are described in the optional Wireless Set-Up Utility
CD.
Summary of Contents for FreeStyle Precision Pro
Page 1: ...Operator s Manual for Healthcare Professionals...
Page 88: ...6 Proficiency Test Proficiency Test Glucose Proficiency Test Ketone Beta Hydroxybutyrate...
Page 98: ...7 Linearity Test...
Page 135: ...9 Review Setup for Wireless Communication Optional Meter Wireless Configuration...
Page 179: ...13 Specifications FreeStyle Precision Pro Meter Symbols Barcode Types...