7-7
Questions? Refer to Page 11-22 For Technical Support Instructions.
Chapter 7: Linearity Test
What You See on the Display
What You Do
Comments
13.
Wait for the meter to analyse the sample
and display the test result.
When using FreeStyle Precision Pro
test strips the countdown will begin at 5
seconds. When using Precision Xceed
Pro test strips the countdown will begin
at 20 seconds.
The meter counts down then displays the
test result. If an error is detected during
the assay, the assay is terminated and no
result is displayed. Once the assay begins,
all key presses are ignored until the assay is
completed.
100 mg/dL
14.
Note the result. If required, scan or
manually enter the comment code and
press
Enter
.
If there is no prompt to enter a
comment code, skip down to step 15.
If the comment code option is enabled in
the control test configuration through the
data management system, the meter will
be enabled to scan or enter a 1 or 2 digit
comment code.
Summary of Contents for FreeStyle Precision Pro
Page 1: ...Operator s Manual for Healthcare Professionals...
Page 88: ...6 Proficiency Test Proficiency Test Glucose Proficiency Test Ketone Beta Hydroxybutyrate...
Page 98: ...7 Linearity Test...
Page 135: ...9 Review Setup for Wireless Communication Optional Meter Wireless Configuration...