8-3
Questions? Refer to Page 11-22 For Technical Support Instructions.
Control Tests
• Set which levels of control solution are required, optional or not used by your facility. The meter can be configured to run
any combination of low, mid and high control solution tests.
• Set whether control solution test results display as numeric or as Pass/Fail.
• Determine whether a comment code is required, optional or not used for control tests that are within the acceptable range
for the facility and the test strip.
• Decide whether a comment code is required, optional or not used for control tests that are outside of the acceptable range
for the facility and the test strip.
QC Interval
• Determine how frequently QC tests need to be run. This can be set to a number of hours or up to three specified times
during the day.
• Decide whether the operator can continue to test (“allow test” or “warn”) or not (“lockout”) when the QC interval is
expired.
• Decide whether the operator can continue to test (“allow test” or “warn”) or not (“lockout”) immediately following strip
port module replacement.
Operator ID
• Decide whether an operator can continue to test (“allow test” or “warn”) or not (“lockout”) when the Operator ID has
expired.
• Decide the number of days before the Operator ID is set to expire that the operator will see a notification on the meter.
Chapter 8: Review Setup
Summary of Contents for FreeStyle Precision Pro
Page 1: ...Operator s Manual for Healthcare Professionals...
Page 88: ...6 Proficiency Test Proficiency Test Glucose Proficiency Test Ketone Beta Hydroxybutyrate...
Page 98: ...7 Linearity Test...
Page 135: ...9 Review Setup for Wireless Communication Optional Meter Wireless Configuration...
Page 179: ...13 Specifications FreeStyle Precision Pro Meter Symbols Barcode Types...