![background image](http://html1.mh-extra.com/html/abbott/freestyle-precision-pro/freestyle-precision-pro_operators-manual_3346701106.webp)
7-8
Questions? Refer to Page 11-22 For Technical Support Instructions.
Chapter 7: Linearity Test
What You See on the Display
What You Do
Comments
100 mg/dL
15.
You can select one of the following
options:
Press
1–New Level
.
Press
2–Same Level
.
Press the
Menu
button to return to the
Menu Mode
.
Press
On/Off
to turn off meter.
Remove the test strip from the meter
and discard it when finished testing.
Follow your facility’s biohazard disposal
policy.
The display shows:
Level
(Test level 1 to 5)
Test
(Replicate 1 to 4 for that level)
Comment
Code
(Number in box)
Test Type
(Glucose)
Date and Time of test
Operator ID
(
o:
prompt)
Note that test date and time alternates with
Operator ID.
If
2
is selected, the
Scan or Enter Strip
Lot
screen will appear. Return to step 8 to
continue linearity testing with the
Same
Level
.
If
1
is selected, the
Select
Level
screen will
appear. Return to step 7 to perform linearity
tests for a
New
Level
.