![background image](http://html1.mh-extra.com/html/abbott/freestyle-precision-pro/freestyle-precision-pro_operators-manual_3346701102.webp)
7-4
Questions? Refer to Page 11-22 For Technical Support Instructions.
What You See on the Display
What You Do
Comments
7.
If the
New
Panel
screen appears you
may either:
Press
1 – ReEnter Kit Lot
.
Press
2 – Replace Panel
.
If the CVC kit lot number that was entered
is different from the previous one, the meter
will prompt you to do one of two actions:
Either re-enter the existing lot number or
replace the existing lot number with the
new lot number.
8.
Select the number of the level of the
next test to be run. If you press
6
for a
New Panel
, the meter will prompt you
to confirm that you wish to replace the
existing panel.
There can be up to 4 replicates per level.
The number in the parentheses indicates
the number of replicates already run for
that level. When a level is full, the number 4
will appear in parentheses, indicating that
all 4 replicates have been completed for this
level. For example
1 - Level 1 (4)
.
Chapter 7: Linearity Test