![background image](http://html1.mh-extra.com/html/abbott/freestyle-precision-pro/freestyle-precision-pro_operators-manual_3346701024.webp)
2-2
Questions? Refer to Page 11-22 For Technical Support Instructions.
Step 3: Slowly move the meter, if needed, so that the green beam is directly over the barcode.
Step 4: The meter will beep in acknowledgement when it accepts the barcode.
When first learning to use the barcode scanner, some precautions should be
taken. It is important that you place the object to
be scanned on a flat surface or hold it by itself. This will prevent other items from being accidentally scanned.
If data has been manually entered, scanning data will erase the manually entered data (unless the Enter key has been selected to
confirm manual entry) and replace it with the scanned data. This entry will be considered scanned. Upon a successful barcode
scan, this system will automatically proceed to the next screen. The scanning beam shuts off in three seconds if nothing is
detected.
Note:
If you hold the Scan button for three seconds, the scanner stops. Reposition scanner and try again.
Optimal distance depends on barcode type.