3-22
Questions? Refer to Page 11-22 For Technical Support Instructions.
Chapter 3: Patient Test
What You See on the Display
What You Do
Comments
100 mg/dL
12.
You can select one of the following
options:
Press
1–Next Patient.
Press
2–Patient History.
Press
Menu
to return to the
Menu Mode
menu.
Press
On/Off
to turn off the meter.
Remove the test strip from the meter
and discard it when finished testing.
Follow your facility’s biohazard disposal
policy.
The display shows:
Patient ID or Name
(
p:
prompt)
Comment Code
(Number in box)
Date and Time of test
Test Type
(Glucose)
If
1
is selected, the
Scan Patient ID
screen
will appear. Return to step 4 to continue
with testing on the next patient.
If
2
is selected, the meter will display the last
test result for this patient as well as allow
you to view previous results.
Refer to the section titled Data Review
for Patient by Patient ID in Chapter 5 for
further information.
IMPORTANT:
If the blood glucose result appears to be inconsistent (lower or higher than expected),
and you have ruled out common errors in technique, there may be a problem with the
test strip. Repeat the test using a new test strip.
Results that are incorrect may have serious medical consequences. Consult the
prescribing physician before making any changes to diabetes medication plans.
Summary of Contents for FreeStyle Precision Pro
Page 1: ...Operator s Manual for Healthcare Professionals...
Page 88: ...6 Proficiency Test Proficiency Test Glucose Proficiency Test Ketone Beta Hydroxybutyrate...
Page 98: ...7 Linearity Test...
Page 135: ...9 Review Setup for Wireless Communication Optional Meter Wireless Configuration...
Page 179: ...13 Specifications FreeStyle Precision Pro Meter Symbols Barcode Types...