3-10
Questions? Refer to Page 11-22 For Technical Support Instructions.
What You See on the Display
What You Do
Comments
6.
Press
Scan
to scan the test strip barcode
or manually enter the test strip lot
number via the keypad, then press
Enter
.
Scanning the barcode identifies the strip
type, calibrates the meter, ensures the
expiry date has not passed, records the strip
lot used, and checks that the lot has been
approved for use by your facility.
7.
The
Sample Type
screen will appear.
You can select one of the following:
Press 1 -
Arterial/Capillary
, or
Press 2 -
Venous
.
Use
Venous
mode only for venous samples
that have been in capped tubes and that
have not been mixed with air. Use
Arterial/
Capillary
mode for all other samples.
If the
Sample Type
screen is not seen,
continue with step 8.
8.
Make sure your hands or gloves are
clean and dry before opening the foil
packet. Open the foil test strip packet at
the notch and tear down to remove the
test strip.
Chapter 3: Patient Test
Summary of Contents for FreeStyle Precision Pro
Page 1: ...Operator s Manual for Healthcare Professionals...
Page 88: ...6 Proficiency Test Proficiency Test Glucose Proficiency Test Ketone Beta Hydroxybutyrate...
Page 98: ...7 Linearity Test...
Page 135: ...9 Review Setup for Wireless Communication Optional Meter Wireless Configuration...
Page 179: ...13 Specifications FreeStyle Precision Pro Meter Symbols Barcode Types...