During Any Test:
Symptom/Error Message
Explanation
How To Respond
The barcode just scanned was not accepted.
Possible reasons are:
• The green scanner beam did not pass across
all the bars of the code.
• The barcode was damaged or poorly
printed.
• The barcode was not of the type specified
for this ID or lot number.
1.
Rescan the barcode or manually enter the
ID or lot number.
2.
Notify the system administrator. The
problem may be the printing of the
barcode or the setup.
The barcode just scanned is from a foil
packet of test strips that has expired.
1.
Discard the expired test strips
2.
Repeat the scan with a new, unexpired
foil packet of test strips.