![ABB FEX300 Operating Instruction Download Page 46](http://html1.mh-extra.com/html/abb/fex300/fex300_operating-instruction_3745924046.webp)
Electrical connections
46 FEX300,
FEX500
OI/FEX300/FEX500-EN
5.3.2 Transmitter
The outer shielding of the cable is attached to the busbar via the clip (3) (from the accessory
bag in the connection area).
The shielding for the signal wires function as a driven shield to transmit the measurement
signal.
The cable is attached to the sensor and transmitter according to the connection diagram.
Fig. 41
1
Terminal cover
2
Busbar (SE)
3
Clip
4
Signal and magnet coil cable
Important (Notice)
The power supply for the optional pre-amplifier is provided via terminals 1S and 2S.
The transmitter automatically detects the sensor and switches to the required supply voltage
on terminals 1S and 2S.
Summary of Contents for FEX300
Page 169: ...Appendix OI FEX300 FEX500 EN FEX300 FEX500 169...
Page 170: ...Appendix 170 FEX300 FEX500 OI FEX300 FEX500 EN...
Page 175: ......