Electrical connections
40 FEX300,
FEX500
OI/FEX300/FEX500-EN
5.1.2 Cable with part number D173D031U01
Prepare both cable ends as shown.
Important (Notice)
Use wire end sleeves.
• Wire end sleeves 0.75 mm
2
(AWG 19), for shielding (1S, 2S)
• Wire end sleeves 0.5 mm
2
(AWG 20), for all other wires
The shields may not touch (signal short circuit).
G01030
70 (2.76)
1
2
3
4
5
6
7
8
9
11 10
X
8 (0.31)
25 (0.98)
"X"
Fig. 34: Flowmeter sensor side, dimensions in mm (inch)
G01031
X
L
2
1
2
3
4
5
6
7
8
9
10
11
8 (0.31)
25 (0.98)
"X"
Fig. 35: Transmitter side, dimensions in mm (inch)
1
Measurement potential 3, green
L2 = 70 (2.76)
2
Signal line E1, violet
L2 = 60 (2.36)
3
Shield
1S
L2
=
60
(2.36)
4
Shield
2S
L2
=
60
(2.36)
5
Signal line, E2, blue
L2 = 60 (2.36)
6
Data line, D2, yellow
L2 = 70 (2.76)
7
Data line, D1, orange
L2 = 70 (2.76)
8
Magnet coil, M2, red
L2 = 90 (3.54)
9
Magnet coil, M1, brown
L2 = 90 (3.54)
10
Groundwire, steel
11
SE clamp