Parameterization
122 FEX300,
FEX500
OI/FEX300/FEX500-EN
Menu / Parameter
Value range
Description
Diagnostics / Fingerprints /
Factory F.P.
1)
Sensor Coil R.
Display only
Elec. Coat. QE1
Elec. Coat, QE2
Elec. Imp. E1-GND
Elec. Imp. E2-GND
Tx check CMR
Tx check 5 m/s
Tx check 10 m/s
The factory fingerprint is created when the device are
calibrated in the factory.
Diagnostics / Fingerprints /
Startup F.P.
1)
Startup F.P.
Create the commissioning fingerprint for the flowmeter
sensor.
Sensor Coil R
Display only
Elec. Coat. QE1
Elec. Coat, QE2
Elec. Imp. E1-GND
Elec. Imp. E2-GND
The commissioning fingerprint is created on site
during commissioning. The values measured for this
fingerprint are indicated here.
Start Tx Startup F.P.
Create the commissioning fingerprint for the
transmitter.
Tx check CMR
Display only
Tx check 5 m/s
Tx check 10 m/s
Diagnostics / Fingerprints /
On demand F.P.
1)
Start On demand F.P.
Creating the manual fingerprint.
Sensor Coil R
Display only
Elec. Coat. QE1
Elec. Coat, QE2
Elec. Imp. E1-GND
Elec. Imp. E2-GND
The manual fingerprint can be created at any time.
The values measured for this fingerprint are indicated
here.
Start Tx Startup F.P.
Create the manual fingerprint for the transmitter.
Tx check CMR
Tx check 5 m/s
Tx check 10 m/s
italics
= Parameter can only be viewed at the "Advanced" password level.
1) Parameter / menu only available for FEP500 / FEH500.