Functional and technical properties - HygienicMaster
164 FEX300,
FEX500
OI/FEX300/FEX500-EN
14 Functional and technical properties - HygienicMaster
Change from one to two columns
14.1 Flowmeter sensor
14.1.1 Protection type according to EN 60529
IP 65, IP 67, NEMA 4X
IP 68 (for external flowmeter sensors only)
14.1.2 Pipeline vibration according to EN 60068-2-6
The following applies to compact devices:
(transmitter mounted directly on the flowmeter sensor)
•
In the 10 ... 58 Hz range with max. 0.15 mm (0.006 inch)
deflection
•
In the 58 ... 150 Hz range with max. 2 g acceleration
The following applies to devices with a separate transmitter:
Transmitter
•
In the 10 ... 58 Hz range with max. 0.15 mm (0.006 inch)
deflection
•
In the 58 ... 150 Hz range with max. 2 g acceleration
Flowmeter sensor
•
In the 10 ... 58 Hz range with max. 0.15 mm (0.006 inch)
deflection
•
In the 58 ... 150 Hz range with max. 2 g acceleration
14.1.3 Installation length
The flange devices comply with the installation lengths specified in
VDI/VDE 2641, ISO 13359, or according to DVGW (process sheet
W420, design WP, ISO 4064 short).
14.1.4 Signal cable (for external transmitters only)
A 5 m (16.4 ft) cable is supplied.
If you require more than 5 m (16.4 ft), a cable can be purchased using
order number D173D027U01.
Alternatively, the cable with order number AD173D031U01 can be
used for transmitters without explosion protection (model FEP321,
FEH321) from DN15 and for transmitters for use in Zone 2 (model
FEP325, FEH325) from DN15.
Preamplifier
Maximum signal cable length between flowmeter sensor and
transmitter:
a) Without preamplifier:
•
Max. 50 m (164 ft) for conductivity
≥
5 µS/cm
A preamplifier is required for cables > 50 m (164 ft).
b) With preamplifier
•
Max. 200 m (656 ft) for conductivity
≥
5 µS/cm
14.1.5 Temperature range
Storage temperature
- 40 ... 70 °C (-40 ... 158 °F)
Min. permissible pressure as a function of fluid
temperature
Lining
Nominal
diameter
P
Operating
mbar abs.
at
T
Operating1)
PFA
DN 3 ... 100
(1/10 ... 4")
0
< 180 °C (356 °F)
PEEK
DN 1 ... 2
(1/25 … 1/12")
0
< 120 °C (248 °F)
1)
For CIP/SIP cleaning, higher temperatures are permitted for limited time periods; refer to
the table titled "Maximum permissible cleaning temperature“.
Max. permissible cleaning temperature
CIP cleaning
Flowmeter
sensor
lining
T
max
T
max
minutes
T
amb
.
Steam cleaning
PFA 150
°C
(302 °F)
60 25
°C
(77 °F)
Fluids
PFA 140
°C
(284 °F)
60 25
°C
(77 °F)
If the ambient temperature is > 25 °C, the difference must be
subtracted from the max. cleaning temperature. T
max
-
∆
°C.
(
∆
°C = T
amb
- 25 °C)
Max. permissible temperature shock
Lining
Max. temp. shock
Temp. diff. in °C
Temp. gradient
°C/min
PFA Any Any
PEEK Any
Any