26
During these tests, the peristaltic pump is stopped (to remove the patient’s heat load from the equation). The
tests that are conducted depend upon the state of the console at the time of the test. The test(s) done is (are)
as follows:
= will be performed
= will not be performed
You will not see any screen messages unless there is an abnormal result in which case the appropriate alarm
state will be called.
Sensor Checks
Figure 4.1. Console Setup Screen
For the console to enter the next state, Standby, the following sensors must be checked and found normal.
If any of these sensor states are not correct you will see the “System Setup” and the problem sensor state will
be highlighted in red (see Figure 4.1 above). Once you rectify the identified problems, the console will progress
to Standby.
Note. Presence of patient temperature probes.
The console can enter Standby without a T1 probe being
connected; however, the console will not enter Run without the primary patient temperature probe (T1) con-
nected to the console.
Coolant Temperature
Heating Test Cooling Test
<10 °C
10 °C - 38 °C
> 38 °C
Air Trap
There is no Start-Up Kit installed or there is a large amount of air in the
Start-Up Kit chamber.
Roller Pump Lid
The clear plastic roller pump lid is not closed properly.
Check Prime Switch
The Prime Switch is being depressed (as when you use it to prime the con-
sole) or has been jammed in the On position.
Coolant
The coolant level is low. Top up the coldwell with coolant.
Table 4.1. Setup Sensor Checks
Содержание IVTM
Страница 2: ......
Страница 4: ...This page intentionally left blank...
Страница 8: ...iv Specifications 101...
Страница 28: ...20 This page intentionally left blank...
Страница 92: ...84 This page intentionally left blank...
Страница 96: ...88 This page intentionally left blank...
Страница 100: ...92 This page intentionally left blank...