34
Hi/Lo Patient Temperature Alarms
The console features two patient temperature alarms: “Hi” and “Lo”. If the alarms are set, the console triggers
an alarm whenever the patient’s temperature is higher than the “Hi” patient temperature alarm value and
whenever the patient’s temperature is lower than the “Lo” patient temperature alarm value. The range of val-
ues for both alarms is 28ºC – 45ºC (82.4ºF – 113.0ºF).
Setting the Alarms
The alarm values are set as follows in either Standby or Run mode. Set the “Lo” and “Hi” patient temperature
alarms to appropriate values.
1. Press the knob and the Main Menu will appear.
2. Select the “Settings” option and press the knob.
3. Select the “Hi/Lo Alarms” option.
4. The “Lo” alarm value will be presented. Turn the knob until the desired value is displayed and press the
knob.
5. The “Hi” alarm value will be presented. Turn the knob until the desired value is displayed and press the
knob.
Hi/Lo Alarms
Allows you to modify the low (Lo) and high (Hi) patient temperature alarms.
Time and Date
Allows you to modify the console time and date.
°C / °F
Allows you to toggle between displaying temperatures in degrees Celsius or
degrees Fahrenheit.
Bath Pre-Set
Allows you to set the console to either Pre-Warm or Pre-Cool during
Standby.
Language
Allows you to select the display language.
Standby Timer
Provides you with a warning alarm after the console has been in Standby for
either 15 or 60 minutes. This alarm can be deactivated.
Exit
Closes the menu.
Table 4.4. Settings Menu
Содержание IVTM
Страница 2: ......
Страница 4: ...This page intentionally left blank...
Страница 8: ...iv Specifications 101...
Страница 28: ...20 This page intentionally left blank...
Страница 92: ...84 This page intentionally left blank...
Страница 96: ...88 This page intentionally left blank...
Страница 100: ...92 This page intentionally left blank...