97
Figure 7.2. Drain Hose and Drain Coupler
5. Place the open end of the drain hose into a suitable container or position it above a floor sink or drain.
6. Push the connector on the drain hose into the drain coupler on the coolant well (refer to Figure 7.3).
Figure 7.3. Push Hose Into Drain Coupler
7. As soon the connector is firmly seated in the coupler, coolant will automatically
begin to flow through the
line.
8. Wait until the coolant flow has stopped. Disconnect the drain line by pressing down on the metal tab on
the coupler. Gently pull out the drain hose.
9. Dispose of the used coolant.
10. Wipe the coolant well dry. Visually inspect the drained coolant well for discolorations or foreign matter.
11. Thoroughly wipe the coolant well with at least 55% isopropyl alcohol or other non-corrosive hospital-
grade disinfectant.
12. Rinse the coolant well with sterile distilled water. Do not rinse with saline.