64
Figure 4.58. Location of the Serial Interface Connector
11. Start the TempTrend CSV program on the laptop computer. See the TempTrend CSV Instructions for Use
(IFU).
12. If a problem occurs during downloading, the screen displays the message “Download error. Please check
external computer.” Check the serial cable connections and the operation of the laptop computer and
choose “Try again.” If repeated failures occur, choose “Cancel.” Repeated failures indicate a problem
with the TempTrend program or the console; contact your ZOLL representative for assistance.
Figure 4.59. Download Error Message
13. When the data download is complete, the screen briefly displays the message “Download Complete.” At
this time, the patient trend data saved from the improper shutdown is deleted.
14. After about two seconds, System Set Up screen displays the message “Set up system now. Press Enter to
proceed.”
15. Disconnect the serial interface cable from the console and the laptop computer.
Downloading of patient trend data is now complete. To set up the console for a new patient, return to the sec-
tion in this chapter titled Preparing the Console for Treatment on page 44.