40
patient’s core temperature. As a result, the console may inappropriately warm the patient. Failure to use a sec-
ond temperature probe may result in patient injury.
First Use Warning – No T2 Probe
For each time the console is powered on:
1. The console will check to see if the Primary (T1) and Secondary (T2) Temperature Probes are present. The
console will enter Standby without probes being present. The console will not enter Run unless the T1
probe is present.
2. When the console is first put into Run, it will check to see which probes are present.
3. If there is no T2 present, the console will ask you to verify that this is intentional. See image below.
Figure 4.18. No T2 Probe Connected
The console remembers the User’s preference.
T2 Probe Disconnection/Reconnection
If you connect a T2 probe at any time during console operation (in Standby or Run), the console will assume
that use of the T2 probe is desired. If the T2 probe is disconnected, the console triggers an alarm – see below.
If the you disconnect a T2 Probe in Standby, the console will not assume that this is intentional. When the con-
sole is placed into Run without a T2 Probe present, the you will see the Warning as described above again.