77
Operating States
The HMIA has the following operating states: Start-up and Run.
Start-up
When the console is powered on, illumination of the Green LED is indication that power has been supplied to
the HMIA. If there is no power to the HMIA the Green LED is not lit and the hospital monitor, if attached to the
HMIA, should detect this as a probe disconnection or temperature probe error.
Once powered, the HMIA immediately establishes its link with the console and performs a series of self tests
including self-calibration – see below.
Run
The HMIA is designed to provide a variable resistance value that is interpreted by YSI-400 compliant hospital
monitors as patient temperature. The patient temperature is obtained via the T1 connection of the console.
Calibration
HMIA performs a self-calibration at Start-up. It will also perform a self-calibration upon reconnection of the RS-
232 port (for example, you might temporarily disconnect the HMIA to allow access for data download). Self-
calibration takes less than half a second. During self-calibration, it may be detected by the hospital monitor as
a disconnected probe. At the end of self-calibration the HMIA will automatically return to Run.
Symbol
Description
Function
RST
Reset Button
This is a reset button. Lightly push the tip of a
paper clip, or similar sized device, into the hole
until you feel a “click” and then withdraw the
paper clip.
Caution LED
This red light indicates that there is a malfunc-
tion in the HMIA.
Power LED
This green light indicates that the HMIA is
receiving power from the console.
RS232
RS-232 Serial port connector
cable
This short extension cable is connected to the
RS-232 data port connector on the back of the
display head.
T1 out
Patient temperature output
connector
This connector accepts the custom ZOLL inter-
face cable.
SYSTEM
Console umbilical cord connec-
tions
One connector is attached to the round, multi-
pin connector on the back of the display head
via a short cable. The other connector mates
with the umbilical cord of the console.
Table 4.7. HMIA Control/Indicators
Содержание IVTM
Страница 2: ......
Страница 4: ...This page intentionally left blank...
Страница 8: ...iv Specifications 101...
Страница 28: ...20 This page intentionally left blank...
Страница 92: ...84 This page intentionally left blank...
Страница 96: ...88 This page intentionally left blank...
Страница 100: ...92 This page intentionally left blank...