Figure 4.31. Check the Following Screen
Installing the Start-Up Kit
Hang the Saline Bag
To install the Start-Up Kit, follow these steps in the indicated order.
1. Obtain an IV bag or bottle of sterile normal saline solution. The bag or bottle must contain no more than
0.5 liter (500 ml) of solution. Hang the saline container on the hook mounted on the rear of the display
head. The container hangs inside the circumference of the handle.
Air Trap
There is no Start-Up Kit installed or there is a large amount of air in the air trap
chamber.
Roller Pump Lid
The clear plastic roller pump lid is not closed properly.
Check Prime Switch
The Prime Switch is being depressed (as when you use it to prime the console) or
has been jammed in the On position.
Coolant
The coolant level is low. Top up the coldwell with coolant.
Table 4.5. Setup Sensor Checks