WARNING. Electric shock risk during cardiac defibrillator discharge.
The console’s protection against the
effect of the discharge of a cardiac defibrillator is partially in the patient temperature probe. To prevent poten-
tial hazards to the patient or operator, the console must be used in conjunction with the approved patient
temperature probes listed in Specifications section of this manual.
Guidance and Manufacturer’s Declaration–
Electromagnetic Emissions
Electromagnetic Immunity Declaration (EID)
The Thermogard XP console is intended for use in the electromagnetic environment specified below. The
customer or user of the console should ensure that it is used in such an environment.
The Thermogard XP console is intended for use in the electromagnetic environment specified below. The
customer or user of the console should ensure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic Environment Guidance
RF Emissions
CISPR 11
Group 1
The Thermogard XP console uses RF energy for its internal function
only. Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby equipment.
RF Emissions CISPR 11
Class A
The console is suitable for use in all establishments other than
domestic establishments and those directly connected to a low-
voltage power supply network that supplies buildings used for
domestic purposes.
Note.
The emissions characteristics of this equipment make it suit-
able for use in industrial areas and hospitals (CISPR 11 Class A). If it
is used in a residential environment (for which CISPR B is normally
required), this equipment might not offer adequate protection to
radiofrequency communication services. You may need to relocate
or re-orient the equipment.
Harmonic Emissions
IEC 61000-3-2
Class A
Voltage Fluctuations /
Flicker Emissions
IEC 61000-3-3
Complies
Table 1.1. Guidance and Manufacturer’s Declaration–Electromagnetic Emissions
Immunity Test
IEC 60601 Test
Level
Compliance Level
Electromagnetic environment --
guidance
Electrostatic dis-
charge (ESD)
IEC 61000-4-2
± 8kV contact
± 15 kV air
± 8kV contact
± 15 kV air
Floors should be wood, concrete, or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be 30% or higher.
Electrical fast transit/
burst
IEC 61000-4-4
±2 kV for power
supply lines
±1kV for input/
output lines
±2 kV for power
supply lines
±1kV for input/out-
put lines
Mains power quality should be that of a
typical commercial or hospital environ-
ment.
Surge
IEC 61000-4-5
±1kV differential
mode
±2kV common
mode
±1kV differential
mode
±2kV common
mode
Mains power quality should be that of a
typical commercial or hospital environ-
ment.
Table 1.2. Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
Содержание IVTM
Страница 2: ......
Страница 4: ...This page intentionally left blank...
Страница 8: ...iv Specifications 101...
Страница 28: ...20 This page intentionally left blank...
Страница 92: ...84 This page intentionally left blank...
Страница 96: ...88 This page intentionally left blank...
Страница 100: ...92 This page intentionally left blank...