5. If necessary, use the vacuum cleaner’s crevice tool to remove dust from the condenser. Gently pass the
crevice tool over all exposed surfaces of the condenser only. Take care not to bend any of the cooling fins
of the condenser. Never use cleaning fluids or water on the condenser.
6. When finished, replace the access door and secure the screws.
7. Inspection and cleaning are complete. The console may be returned to service.
Temperature Accuracy Test
Required Tools and Materials
You will need a calibrated temperature source (e.g., Fogg System Co. model TP-400 or equivalent).
Procedure
To perform a temperature accuracy test, follow these steps in the indicated order:
1. Set the calibrated temperature source to exactly 37º C and plug it into connector T1.
2. Start the console and proceed to the Standby screen.
3. Observe the patient temperature displayed on the screen. It should indicate 37º C ± 0.2º C.
4. If the displayed temperature is above or below the indicated range, contact your ZOLL service
representative.
5. If the displayed temperature was within the indicated range, the test is complete.
Clean Console and Condensate Pan
Required Tools and Materials
You will need a soft, lint-free cloth and a solution of mild detergent and water to perform these tasks.
Procedure
To clean the console, follow these steps in the indicated order:
1. Ensure that the console has been switched off and the power cord has been unplugged.
2. Clean the exterior of the console using a soft cloth dampened with a mild detergent and water mixture.
Never use solvents or abrasive cleaners on the console. Avoid vigorous scrubbing, especially on the front
surface of the display.
Note. Console disinfection.
To disinfect the console, ZOLL recommends PDI
®
Super SANI-CLOTH
®
ger-
micidal disposable wipes (EPA Reg. No: 9480-4). An equivalent product approved by local authorities may
be used. It must include at least 55% isopropyl alcohol or other approved non-corrosive hospital-grade
disinfectant. ZOLL highly recommends testing the disinfectant on a small area on the outer surface of the
console prior to wiping all surfaces of the console. Follow directions provided by the manufacturer of the
disinfectant. The hospital determines the effectiveness of the disinfectant.
3. Wipe down the outside surfaces with a water-dampened cloth to remove remaining spots or residue.
4. Slide out and remove the condensate collection pan from under the front of the console (see Figure 7.1).
Содержание IVTM
Страница 2: ......
Страница 4: ...This page intentionally left blank...
Страница 8: ...iv Specifications 101...
Страница 28: ...20 This page intentionally left blank...
Страница 92: ...84 This page intentionally left blank...
Страница 96: ...88 This page intentionally left blank...
Страница 100: ...92 This page intentionally left blank...