60
Do not keep replenishing saline that is being rapidly depleted–a problem exists that must be immediately rem-
edied.
For additional warnings and potential risks, see Operating Life on page 11.
Setup - Variations
Setup Sequence
The Setup sequence varies depending upon the following factors:
1. The time from the last power down.
2. Whether or not you used the “End Procedure” menu option when you terminated the previous case.
Time From Last Power Down
If the console detects that it is less than 3 minutes since the last time it was powered down it will dispense with
some of its self tests depending upon the state of the cooling engine at the time of power down. The idea
being that, with such a short time, it is more likely than not that the power down was either unintentional (you
tripped over the power cord) or an intentioned brief adjustment that required the console to be powered off
(you moved the console to a new position).
The console will give you the option to use the last programmed settings. If you elect to do this you may still
use the menu to change console settings in Standby.
Figure 4.52. Operate at Current Settings
If it has been less than 3 minutes since the last power down, when you power on the console, the screen
above is displayed.
1. If you select “Yes” and press the knob, you will proceed directly to Standby once the console self tests
have completed.
2. If you select “No” and press the knob, you will continue to be presented with screens that allow you to
select operating parameters for the console prior to entering Standby starting at Figure 4.55– see below.
Содержание IVTM
Страница 2: ......
Страница 4: ...This page intentionally left blank...
Страница 8: ...iv Specifications 101...
Страница 28: ...20 This page intentionally left blank...
Страница 92: ...84 This page intentionally left blank...
Страница 96: ...88 This page intentionally left blank...
Страница 100: ...92 This page intentionally left blank...