57
Figure 4.46. Route the Tubing Out of the Console
21. Lift the saline container off the hook and slip the insulating jacket around the container. Carefully close
the hook-and-loop fasteners at the top and bottom of the jacket. Rehang the container on the hook.
22. When the self tests are completed and the Start-Up Kit is loaded and primed, the console will enter
Standby. The console is ready to be connected to the patient.
Figure 4.47. Standby Screen – T1 Probe Not Connected
Connecting the Patient to the Console
WARNING. Verify console function first.
Ensure proper functioning of the console and initiate pre-cool or
pre-warm (if applicable) prior to placing the catheter in the patient.
When the console has been prepared as directed in the preceding sections, it may be moved to the patient’s
bedside and connected to the patient. Follow these steps in the indicated order.
1. Position the console near the patient’s bed. It must be close enough so that the temperature probe cables
and the tubing can conveniently reach the patient. Route the cables and tubing safely.