  TECHNICAL BULLETIN TB-6568 

Description

The EMIT Ionization Test Kit allows the 

50597

 Digital 

Static Field Meter to be used to measure the offset 

voltage (balance) and charge decay of ionization 

equipment.  The Test Kit also includes a Charger used to 

place a ±1000V charge on the 

50567

 Conductive Plate, 

making it possible to also measure the discharge times 

of air ionization equipment per ANSI/ESD SP3.3 Periodic 

Verification of Air Ionizers.  The 

50598

 Ionization Test Kit 

includes the 

50597

 Digital Static Field Meter, providing a 

highly portable and cost effective means of verifying the 

performance of a wide variety of ionization equipment.

Note:

 The 

50597

 Digital Static Field Meter is designed to 

operate only with the 

50598

 Ionization Test Kit.  It is not 

compatible with other brands.
Although not as accurate, the EMIT Ionization Test Kit 

has been designed to make measurements that 

correspond to those made by using a charged plate 

analyzer and ANSI/ESD S3.1.  The Ionization Test Kit 

provides convenience and portability to test per ANSI/

ESD SP3.3 Periodic Verification of Air Ionizers or 

Compliance Verification ESD TR53.  We recommend 

EMIT’s 

50555

 / 

50561

 Charged Plate Analyzer if precise 

measurements are required.
The Ionization Test Kit includes a slide-on isolated 

Conductive Plate, a ±1000 volt Charger and a durable 

thermoplastic carrying case with custom cut-outs for all 

of the above components along with the model 

50597

 

Digital Static Field Meter.

Charged insulators in the ESD protected area can 

adversely impact quality, productivity, and reliability.
“When any object becomes electrostatically charged, 

there is an electrostatic field associated with that charge.  

If an ESDS (ESD sensitive) device is placed in that 

electrostatic field, a voltage may be induced on the 

device.  If the device is then momentarily grounded, a 

transfer of charge from the device occurs as a CDM 

(Charged Device Model) event.  If the device is removed 

from the region of the electrostatic field and grounded 

again, a second CDM event will occur as charge (of 

opposite polarity from the first event) is transferred from 

the device.”  (ESD Handbook ESD TR20.20 section 

2.7.5 Field Induced Discharges) 
Compliance verification should include periodic checks 

with a static field meter to determine if high 

charging material is present in the ESD protected area.  

All packaging and other materials that may be 

electrostatic generative to 2,000 volts must be kept a 

minimum of 12" from ESD sensitive items at all times.  It 

is proper to rub an item and measure that it can charge.
“In order to mitigate field-induced CDM (Charged Device 

Model) damage, the ESD program shall include a plan 

for the handling of process-required insulators.  If the 

field exceeds 2,000 volts/inch, steps shall be taken to 

either:

A. Separate the insulator from the ESD-sensitive device

  by a distance of 30 cm (12 inches); or

B. Use ionization or other charge mitigating techniques 

  to neutralize the charge.”  (ANSI/ESDS20.20 section 

  8.3)
Other steps that can be taken are to remove the item 

from the ESD protected area, periodically coat with a 

topical antistat, or replace with a static control protective 

version of the item.

Packaging

1  Digital Static Field Meter

1  Conductive Plate

1  Charger

2  9V Alkaline Batteries

1  Data Output Cord

1  Carrying Case

1  Certificate of Calibration

Figure 1. EMIT 

50598

 Ionization Test Kit

© 2010 DESCO INDUSTRIES INC.

Employee Owned

Summary of Contents for TB-6568

Page 1: ...nsulators in the ESD protected area can adversely impact quality productivity and reliability When any object becomes electrostatically charged there is an electrostatic field associated with that charge If an ESDS ESD sensitive device is placed in that electrostatic field a voltage may be induced on the device If the device is then momentarily grounded a transfer of charge from the device occurs ...

Page 2: ...nt operation B RANGE ZERO Button Press to select the measurement range Press and hold to zero the Meter C Analog Output Jack A low voltage replica of the measured voltage is provided at this output The voltage at this output is 1 1000th 2 kV range or 1 10 000 20 kV range of the measured voltage D POWER Button Press to turn the unit ON and OFF FRONT VIEW BACK VIEW 1 2 3 4 Figure 3 Charger features ...

Page 3: ...rate measurements INSTALLING THE ISOLATED PLATE ASSEMBLY The Digital Static Field Meter s case has two slots along its sides The top slot is closest to the face of the instrument Slide down the tabs of the Conductive Plate plate into the top slot of the Meter s case as far as they go see Figure 4 BATTERY CHECK The battery should be replaced when BAT is indicated on the display Always replace the b...

Page 4: ...h that powers the unit Holding the switch forward backward supplies power to the output terminals POLARITY SELECTION The top of the rocker switch is labeled and the bottom is labeled To provide a POSITIVE voltage output touch the plate located underneath the charger and press the switch forward at the same time To provide a NEGATIVE voltage output touch the plate located underneath the charger and...

Page 5: ...isopropyl alcohol Then allow the instrument to air dry for several hours Specifications DIGITAL STATIC FIELD METER Measurement Range switch selectable Low Range 0 to 1 99 kV inch High Range 0 to 19 99 kV inch Measurement Accuracy Voltage Monitor Output 5 of reading 10 mV Voltage Display 5 of reading 2 counts Measurement Stability 10 counts Automatic Shutoff Unit will shut off after 20 minutes afte...

Page 6: ... Return Material Authorization RMA and proper shipping instructions and address EMIT will quote repair charges necessary to bring your unit up to factory standards Warranty Exclusions THE FOREGOING EXPRESS WARRANTY IS MADE IN LIEU OF ALL OTHER PRODUCT WARRANTIES EXPRESSED AND IMPLIED INCLUDING MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE WHICH ARE SPECIFICALLY DISCLAIMED The express warran...