6. CALIBRATING THE MINIGAS MK5
b.
Aspirator plate assembly (part no. 325-9374-00) (used
as a calibration hood).
c.
2mm A/F hexagonal wrench (Allen key).
d.
Gas tubing of suitable size and length for connecting
the gas cylinder to the calibration hood.
Note: Tubing lengths should be kept to an absolute minimum
to reduce the effects of gas adsorption (see
Section 6.3.1) and, in the case of chlorine test gas, to
avoid the need for extensive purging (see Section 6.3.2).
Zellweger Analytics can supply calibration kits containing the
calibration hood, gas tubing and the appropriate low flow
regulator - refer to Section 8 for part numbers.
6.3
PREPARATION FOR CALIBRATION
WARNING
The test gases are vented to atmosphere during the
following purging and calibration procedures. Ensure
adequate ventilation or use a fume cupboard.
6.3.1
General
Calibration must take place in a clean, fresh air environment at
an ambient temperature of between 20° and 25°C (68° to 77°F).
Smoking or the use of butane lighters, solvents etc nearby can
cause faulty, and hence potentially dangerous, calibration.
Ensure that the Minigas Mk5 is fitted with either a fully charged
battery pack or a new set of dry cells.
6.3.2
Adsorption of the Test Gas
Instead of passing freely along the gas tubing, molecules of
the test gas can stick onto the tubing wall, thereby reducing