5. FIELD SETTABLE ALARM MODE
Note: Access to the internal calibration button is obtained by inserting
the long shaft of the Allen key through the screw hole.
(3)
Enter Field Settable Alarm mode by holding the
calibration button depressed while switching on the
Minigas Mk5, ie while pressing and releasing the I/O
button. The AdJ AL. screen is then displayed.
(4)
Select No to return to the normal start-up sequence, or
Yes to continue with the alarm setting adjustments.
5.2.3
Oxygen Alarms
(1)
Obtain the AdJ AL. screen (see Section 5.2.2) and select
Yes to obtain the AdJ -O- screen.
(2)
Select Yes to obtain the -O- Lo screen indicating the
current oxygen instantaneous low alarm trigger level.