3
MAN0585.P65
Issue 01 - 11/99
2987M5023
WARNINGS AND CAUTIONS
WARNING
The Minigas Mk5 is to be serviced, and the battery charged, only in a safe
area.
Substitution of components may impair the intrinsic safety of the
Minigas Mk5. Replacement parts must be approved by Zellweger Analytics.
This applies to all parts, including sensors, components and accessories.
CAUTION
The Minigas Mk5 must be serviced only by qualified personnel trained
by Zellweger Analytics or by a Zellweger Analytics appointed agent.
This Reference Manual must be read and fully understood before
servicing the Minigas Mk5.
Note:
Failure to observe and abide by the above Warnings and Cautions
may render the intrinsic safety approval void and may remove any
right of claim against Zellweger relating to product liability or
consequential damage to any third party.
CAUTION
The following gases or substances may contaminate the explosive gas
sensor (pellistor) in the Minigas Mk5:
Silicones
Halogens (>100 ppm)
Halogenated hydrocarbons
Phosphorous-containing compounds
Sulphur containing compounds (>100 ppm)
Volatile organometallics
If the Minigas Mk5 has been exposed to any of the above contaminating gases
or substances, carry out the calibration check described in Chapter 6.