2. USING THE MINIGAS MK5
Either of the following warning screens, accompanied by the
fault alarm bleep, stops the start-up sequence but can be
overridden by applying a medium or long press to the I/O button.
Log Full
If the Minigas Mk5 is configured to have a fixed log and the log
store is full, LOG Full appears in the upper two banks. This
warning indicates that the instrument will not log any more data
until the current data has been downloaded (see Section 3.6).
Clock Not Set
It is important that the events recorded in the Minigas Mk5
datalog should relate to absolute dates and times. To achieve
this, datalogging instruments contain a clock/calendar device
which is supported by a small back-up battery. If this back-up
power is interrupted, a warning is given by SEt rtc appearing in
the upper two banks. Continuing to use the Minigas Mk5
after overriding this warning will result in incorrect dates
and times being recorded in the data log.
Use the LogView Datalogging Software to reset the clock/
calendar.
2.4.4
If Countdown Stopwatch Alarm is Configured
2.4.4.1 Timer Duration Screen
If the Countdown Stopwatch Alarm option is configured, an
additional screen is displayed at the end of the normal start-up
sequence. This screen (SEt dur) shows the timer duration that
has been set into the Minigas Mk5. For example, the following
illustration shows a timer duration of 1 hour 15 minutes.