4.6
The Light Manager
It is the function of the Light Manager temporarily to create optimum illumination settings for the various
contrast methods and magnifications used and to make these settings reproducible to the user by
providing the possibility to store them permanently.
The Light Manager has three operating modes: OFF, CLASSIC, SMART. The scope of functions available in
the individual modes depends on various optional stand components.
The Light Manager is available for the transmitted-light contrast methods (brightfield, phase contrast,
DIC, darkfield, polarization), for the reflected-light contrast methods (brightfield, darkfield, DIC,
polarization) and for reflected light fluorescence. When working in reflected light, the motorized stop
slider, the neutral-density double-filter wheel and the motorized fluorescence attenuator, as far as
available, are also included in the Light Manager functionality.
The following table shows which operating modes of the Light Manager can be used for the individual
stand models and the contrast methods mentioned.
Axio Imager
Stand
A2
LED
A2
A2m
D2
D2m
M2
M2m
Z2
Z2m
OFF
temporary
+
+
+
+
+
-
-
-
-
OFF
permanent
-
-
-
-
-
+
+
+
+
CLASSIC
TL
+
+
+
+
+
+
+
+
+
CLASSIC
RL (MAT)
+
+
+
+
+
+
+
+
+
CLASSIC
FL
(BioMed)
-
-
-
-
-
-
-
+
+
SMART
TL
-
-
-
-
-
+
+
+
+
SMART
RL (MAT)
-
-
-
-
-
+
+
+
+
SMART
FL
(BioMed)
-
-
-
-
-
-
-
-
-