FIRST-TIME SET-UP
Axio Imager 2
Setting stops for the adjustment range of the stage carrier dovetail guide
48
430000-7544-001
01/2016
3.18
Setting stops for the adjustment
range of the stage carrier dovetail
guide
Lower stop screw (Fig. 30/
5
):
•
When employing the reflected-light / trans-
mitted-light stage carrier with condenser carrier
430707-0000-000, use a slot-head screwdriver
to screw the lower stop screw into the upper
tapped hole (Fig. 30/
3
).
•
When employing the reflected-light or
reflected-light/transmitted-light stage carrier
430704-9901-000 or 430701-9901-000, screw
the lower stop screw into the lower tapped hole
(Fig. 30/
4
).
Upper stop screw (Fig. 30/
6
):
•
When an 18-mm thick microscope stage is
used, screw the upper stop screw into the
tapped hole marked with number "18"
(Fig. 30/
1
).
•
When a 20-mm thick microscope stage is used,
screw the upper stop screw into the tapped
hole marked with number "20" (Fig. 30/
2
).
The thickness of the stage is shown right next to the order number on a sticker attached to the
microscope stage.
If there is no sticker on the microscope stage, the stop screw is to be set to 18 mm to maintain
focusability.
Fig. 30
Setting dovetail guide stops