Axio Imager 2
OPERATION
ZEISS
Illumination and contrast methods
01/2016
430000-7544-001
177
−
After inserting the compensator
λ
and rotating the specimen by 90
°
(2 click stops), the interference
color remains white. It is then, however, a "higher-order white" and thus the path difference is > 5
λ
.
The appropriate compensator is:
Tilting compensator K 0-30
λ
(Accessory 000000-1115-698)
−
A dark gray as interference color suggests very small path differences (
λ
/10 or 54.6 nm).
•
Push the compensator into the slot as far as it will go.
Use the enclosed operating instructions for preparing and conducting the measurement.