FIRST-TIME SET-UP
Axio Imager 2
Mounting the TFT display on the motorized stand
54
430000-7544-001
01/2016
3.23.2
Deactivating the high-performance
focus
•
Use the Z drive to move the microscope stage
into the lowest position.
•
Using the AF 3 ball-head screwdriver (Fig. 37/
1
)
release the two locking pins (Fig. 37/
2
) on the
right
and
left of the stand (Fig. 37/
3
) by turning
them
90 degrees anticlockwise
4
).
The head of the locking pin is now no longer
visible, and the high-performance focus is
inactive.
•
Use the Z drive to move the microscope stage
into the working position.
3.24
Mounting the TFT display on the
motorized stand
When mounting the TFT display, the
microscope must be switched off.
•
Mount the TFT display (Fig. 38/
2
) on the right
side of the motorized stand (Fig. 38/
1
) using the
three screws (Fig. 38/
3
).
During the mounting, the stand and TFT display
are automatically connected electrically via the
plug contact.
Fig. 37
Deactivating the high-performance
focus
Fig. 38
Mounting the TFT display