ZEISS
OPERATION
Axio Imager 2
Operating the motorized microscope via the touchscreen of the TFT display
124
430000-7544-001
01/2016
For XY and Z, the two pairs of
Set Zero
buttons function as follows:
man
Manual
zero setting, i.e. the current position is defined as the zero point and the display set to
zero.
auto
Automatic
zero setting, i.e. the stage moves into the end position that was defined as the zero
point. The display is then set to zero.
Before the stage moves into the bottom Z end position, the following popup window appears:
"Caution! Remove specimen before stage moves into end position!" Confirm the message with
OK
if no specimen is on the stage, or with
Cancel
to stop stage movement.
Save Position
Using the
Save Position
button, the operator can
define coordinate positions for the five lower
position buttons as follows:
•
Move to the desired XYZ position.
•
If the Z value is to be stored, activate the
incl.
Z
check box.
•
Press the
Save Position
button. The
Current
positions save as
pop-up window opens
There are five buttons, Pos.1 ... Pos.5, in the
window. If a button is occupied, the XYZ data
appear as labels, otherwise its number is shown.
•
Save the current position by pressing a position
button. If coordinate data have already been
assigned to this button, a safety query appears
asking you if you want to overwrite the stored
data.
•
Close the window using the
Cancel
button.
To delete data, press the
Delete
button, select the
position button and confirm the safety query with
Yes
.
The positions will only be saved
temporarily, i.e. they will be available
only as long as the microscope is
switched on.
Moving to saved position
There are five buttons in the bottom
Position
field. Press a button to move to the saved
coordinate position. How to save coordinate
positions is described above in
Save position
.
Fig. 123
Microscope -> XYZ-> Save Position
Fig. 124
Microscope -> XYZ -> Move to
Position