Axio Imager 2
OPERATION
ZEISS
Switching the HBO 100 on/off
01/2016
430000-7544-001
95
4.4
Switching the HBO 100 on/off
•
The HBO 100 illuminator, which is used in place of the HAL 100 illuminator for fluorescence contrast
examinations, is switched on and off using the on/off switch (Fig. 85/
2
) of the HBO 100 transformer.
•
After switching the illuminator off, allow it to cool down for approx. 15 minutes before switching it on
again. Failure to do so will unnecessarily shorten the service life of the mercury vapor short-arc lamp.
4.5
Binocular phototube 30°/25 mot.
with two camera ports (2TV tube
mot.)
The 2TV tube mot. (Fig. 86) is available for all
Axio Imager stands of type .M2 / .M2m and .Z2 /
.Z2m.
It is mounted on the coupler plate like any other
binocular tubes and connected via CAN bus to the
stand. To do so, connect the cable to one of the
three available CAN connecting sockets (Fig. 59).
The 2TV tube has two TV ports (Fig. 86/
2
).
The front TV port is adjustable in the X, Y and Z
axis and to the rear (fixed) TV port.
The tube is configured via the TFT display or the
MTB 2011.
LEDs on the tube display (Fig. 86/
3
, Fig. 87)
indicate the selected light path setting.
If you find the illumination of the display a
hindrance, you can switch it off by pushing the
eyepiece shutter button (Fig. 88/
1
) for three
seconds.
Fig. 86
Binocular phototube 30°/25 mot.
with two camera ports (2TV tube
mot.)
Fig. 87
Tube display