OPERATION
Axio Imager 2
The Light Manager
100
430000-7544-001
01/2016
4.6.4
Light Manager of the coded microscope
The Light Manager of the manual microscope controls the light intensity for specimen observation via the
lamp voltage. The functioning depends on the Light Control used:
Light control, manual
Light control, motorized
Lamp voltage saved for each
−
Objective nosepiece position (coded)
Lamp voltage saved for each
−
Objective nosepiece position (coded)
−
Reflector turret position (coded)
−
Tube lens turret position (motorized)
On manual stands, the CLASSIC mode of the Light Manager is active. In this mode, no illumination values
are computed for the range of objectives used. The operator adjusts the brightness for each objective
individually. The values are stored permanently by pressing the
LM-Set
button on the rear of microscope.
This must be done for each individual objective.
When an objective is swung into the light path for which the light intensity has already been set, the
system will automatically set the corresponding stored value.
The light intensity settings of the halogen lamp or LED illuminator are stored separately for transmitted
and reflected light. If the Light Manager is used for reflected-light contrast methods, be sure to pre-set
the voltage selector on the back right of the stand to reflected light. Otherwise, the Light Manager will
assume that the stand is configured for fluorescence illumination. If a motorized neutral-density double
filter wheel has been configured in reflected light, it will be set to 100 % transmission and is thus
inactive.
On the Axio Imager.A2, .A2m, .D2, .D2m manual stands, the Light Manager can be deactivated
temporarily. To do so, keep the RL reflected-light shutter on / off button (Fig. 76/
11
) depressed
while switching on the microscope.
4.6.5
Light Manager and Dazzle Protection control on manual stands
•
Keep the
RL
button depressed while switching on the instrument:
->
Light Manager (LM) & Dazzle Protection (DP)
are permanently deactivated / activated