13-15
IM 703155-01E
Rear Panel Auxiliary Input/Output
13
Sequence Trigger-External Sampling Clock Input
CAUTION
Only input signals that meet the specifications below. Otherwise, undesirable
signal such as excessive voltage may damage the VB8300.
The sampling clock for sequence trigger (see section 7.2) can be input externally. Apply
a signal with a rising pulse width of 100 ns or greater.
Sequence Trigger-External Sampling Clock Input Terminal (SEQ TRIG SMPLG IN)
Specifications
Item
Specifications
Connector type
BNC
Input level
3.3 V CMOS
Input impedance
10 k
Ω
or more
Allowable input level range 0 to 5.5 V
Note
For a description of its use, see section 7.2.
13.7 Sequence Trigger-External Trigger Source and External Sampling Clock (Option)