13-13
IM 703155-01E
Rear Panel Auxiliary Input/Output
13
13.7 Sequence Trigger-External Trigger Source and
External Sampling Clock (Option)
Sequence Trigger-External Trigger Source Input
CAUTION
Only input signals that meet the specifications below. Otherwise, undesirable
signal such as excessive voltage may damage the VB8300.
The trigger source for sequence trigger (see section 7.2) can be input externally. Apply
a signal with a rising pulse width of 100 ns or greater.
Sequence Trigger-External Trigger Source Input Terminal (SEQ TRIG TRIG IN)
Specifications
Item
Specifications
Connector type
BNC
Input level
3.3 V CMOS
Input impedance
10 k
Ω
or more
Allowable input level range
0 to 5.5 V
Note
For a description of its use, see section 7.2.