13-6
IM 703155-01E
10-MHz Reference Signal Output
CAUTION
Do not apply external voltage to the 10 MHz REF OUT terminal. If you do, the
VB8300 may malfunction.
The 10-MHz reference signal is output externally. When using the internal 10-MHz
reference signal, the internal reference signal is output. When using an external 10-MHz
reference signal, the external reference signal is output. For a description of the internal/
external reference signal selection, see section 5.1.
10-MHz Reference Signal Output Terminal (10 MHz REF OUT)
Specifications
Item
Specifications
Connector type
BNC
Output frequency
10 MHz
±
1 ppm
Output level
TTL
Output impedance
Approx. 50
Ω
Output Circuit
5 V CMOS
50
Ω
Note
A synchronized operation is possible using the 10-MHz reference signal input/output. For
details on the synchronized operation of multiple units, see section 6.3.
13.3 10-MHz Reference Signal Input/Output