IM 701240-02E
Recording T-Y (Time-Amplitude) Waveforms
Setting the Display Format to 1 Zone
Section 8.1, "Changing the Display Format" in the user's manual
By default, the screen is divided into 4 zones. Because we are only observing CH1, we set the display to 1 zone so that the
waveform amplitude can be recorded as large as possible.
1
Press
DISPLAY
.
2
Display the selection menu and select 1
Zone.
A single screen is displayed.
DISPLAY
Changing the Voltage Range to 5.00 V
Section 5.3, "Setting the Voltage Range" in the user's manual
Here, we change the voltage range so that the entire sine wave with an amplitude of approximately 5 V is displayed and
recorded as large as possible.
1
Turn the
RANGE
knob clockwise to set the voltage range to 5.00 V.
Voltage range setting of CH1
Note
If you want to change the voltage range
(measurement range) of a channel other than CH1,
press , select the desired channel on the
channel selection menu, and turn the RANGE knob.
Ground level mark
Top section:
Voltage range before the change (voltage range that can be
changed before the voltage range is changed*)
Bottom section:
New voltage (voltage range that can be displayed when the
waveform acquisition is started)
* Up to this point, the steps were carried out without the
waveform acquisition in progress. Because the
acquisition is not in progress, the voltage range for
acquiring waveforms is not updated even if the voltage
range setting is changed.
The voltage range before the change is displayed at the
top section as Span, and the new voltage range is
displayed in the bottom section as Range.
RANGE
CH